Datascope Getinge Recalls Cardiosave Hybrid, Cardiosave Rescue, CS300 and CS100/100i Intra-Aortic Balloon Pumps

Published: 07/24/2019

Sources:
  • Food and Drug Administration
Relevant To:
  • All Healthcare Organizations

Maquet/Datascope is recalling all Intra-Aortic Balloon Pumps (IABPs) due to reports of the IABP batteries failing to hold a charge, stopping unexpectedly, and having a shortened run-time which may cause the device to stop working when being operated by battery only. This recall is being conducted to ensure that all IABP users and servicers follow each device's Operating Instructions Manual for recommendations on usage, charging, maintenance and storage of the batteries, as battery run times and